Artificial breast fatigue testing machine
Product description
Executive standard: yy0647-2008 Appendix E requirements
Product introduction:
It is composed of a fixed plate and a movable plate, the latter is connected with the motor through a connecting arm, so as to introduce alternating motion. The movable plate also includes an adjusting mechanism, which can conveniently change the distance between the movable plate and the fixed plate, so that the implant can be compressed as required. The total length of the movable plate travel should be 40mm, equivalent to moving 20 mm from the middle starting position to each direction, and adjusting the motor speed to generate 200 cycles per minute (3.3hz).
The amplitude is 40mm ± 0.3, the height is adjustable, and the reading accuracy is 0.01mm
The roughness of contact surface is less than 0.1 μ M
Height difference of four corners < 0.2mm